CHEMISTRY ANALYZER
FULLY-AUTOMATED
SAMPLE/REAGENT POSITION
- 40pcs sample position
- Reagent refrigerated function
- 40pcs reagent position
REACTION CUVETTES
- 60pcs reusable plastic cuvettes
- Automatic blank testing for cuvette
- Colorimetry system in the pattern of light splitting behind cuvette
- Dismountable, easily replacement
SAMPLE/REAGENT/MIXING PROBE
- Automatic liquid level detect
- Collision protection function
- Probe internal/external washing
- Independent mixing probe
7-STEP AUTO CLEANING SYSTEM
- Automatic 7-step washing
- Hi-tech washing station to ensure accurate results
- With low water consumption: 1-2L/H
SYSTEM FUNCTIONS
- Analysis Methods: end-point, kinetic, two-point, double-reagent, double-wavelength, multi-standard etc, open to various reagents
SAMPLE/REAGENTS HANDLING
- Sample volume: 1.6-50uL, 0.1uL step
- Sample Dilution: automatic/Manual pre-dilution, with dilution ratio up to 1:100
- Reagent Position: 40pcs
- Reagent Volume: 10-500uL, 0.5uL step
TC200
FULLY AUTOMATED CHEMISTRY ANALYZER
OVERALL PERFORMANCE
Equipment Type
Analysis Rate
Test Principle
Analysis Method
SAMPLE SYSTEM
Sample Tray
Sample Volume
Sample Probe
Sample Containers
REAGENT SYSTEM
Reagent Tray
Reagent Volume
Reagent Probe
Reagent Refrigeration
Fully automatic discrete, STAT priority.
Constant speed 300T/H. (without ISE)
Colorimetry, turbidimetry.
End-point 2 kinetics, fixed-time etc
single/double wavelength, linear and non-linear calibration.
49 Sample positions, random access of all samples
2-50ul, 0.1ul increment.
Liquid level detection, the remaining detection, collision protection etc.
Sample cups, Micro tube. Original blood-collection tube, etc.
56 reagent positions
R1:25-300ul, R2:10-15ul, 0.1ul increment
Liquid level detection, the remaining detection, collision protection etc.
4-15°C, independent power supply, 24-hour non-stop refrigeration.
NT 300
FULLY AUTOMATED CHEMISTRY ANALYZER
FEATURES
- Fully automatically, discrete, random STAT function
- Online test items: at most 40 testing items
- Throughput: 260 tests/hour
- Analysis Method: end-point, kinetic, two-point, double
- Reagents, double-wavelength, multi-standard etc, open to various reagents
- Data Processing: memory up to 200,000 patient's data
SAMPLE/REAGENTS HANDLING
-Sample Position: 40pcs (Include standard, QC, STAT positions)
- Sample volume: 2-50uL, 0.1 uL step
- Sample Probe Cleaning: automatic internal & external washing
- Sample Dilution: automatic/Manual pre-dilution, with dilution ratio up to 1:100
- Reagent Position: 40pcs (with refirgerated function, 2°C-8°C)
- Reagent Probe: automatic liquid detect, with collision protection function
- Reagent Probe Cleaning: automatic internal & external washing
REACTION SYSTEM
- Reaction Cuvette: 90ocs made of reusable plastic
- Opical Length of Cuvette: 6mm
- Reaction Volume: 200-500uL
- Reaction Time: 8-10 minutes
- Reaction Temperature: 37°c-0.1°c
- Mixing System: Independent mixing stirrer
OPTICAL SYSTEM
- Light Source: halogen-tungsten lamp
QC
- Various QC managements, insert QC randomly; save, display & print QC chart; define different QC in advance
TC6030
FULLY AUTOMATED CHEMISTRY ANALYZER
FEATURES
- Fully automatically, discrete, random STAT function
- Online test items: at most 80 testing items
- Throughput: 400 tests/hour
- Analysis Method: end-point, kinetic, two-point, double
- Reagents, double-wavelength, multi-standard etc, open to various reagents
- Data Processing: memory up to 2 million patient's data
SAMPLE/REAGENTS HANDLING
-Sample Position: 93pcs (Include standard, QC, STAT positions)
- Sample volume: 2-50uL, 0.1 uL step
- Sample Probe Cleaning: automatic internal & external washing
- Sample Dilution: automatic/Manual pre-dilution, with dilution ratio up to 1:100
- Reagent Position: 40pcs (with refrigerated function, 2°C-8°C)
- Reagent Probe: automatic liquid detect, with collision protection function
- Reagent Probe Cleaning: automatic internal & external washing
REACTION SYSTEM
- Reaction Cuvette: 90pcs made of reusable plastic
- Optical Length of Cuvette: 6mm
- Reaction Volume: 200-500uL
- Reaction Time: 8-14 minutes
- Reaction Temperature: 37°c-0.1°c
- Mixing System: two Independent mixing stirrer
OPTICAL SYSTEM
- Light Source: halogen-tungsten lamp
- Wavelength: 340-750nm, 12 wavelengths, precision 2nm
- Spectrophotometry: rear spectroophotometry
- Test Range: 0-4.0 Abs
- Resolution: 0.001Abs
TC6090
FULLY AUTOMATED CHEMISTRY ANALYZER
SEMI-AUTOMATED
Test mode: Kinetic, End point, Two point, Absorbance
Wavelength: 7 standard filters, 340,405,492,510,546,578,630nm, 1 free position
Wavelength precision: ±2nm, width ≤ 10nm
Absorbance range: 0~4.500 Abs
Absorbance precision: outside 0.0001 Abs, inside 0.00001 Abs
Carryover: <1%
Repeatability CV: <1%
Incubator: 37°C, 30°C, 25°C, and others, Precision: ±0.1°C
Sample Volume: 0~3000ul (500ul recommend)
Data Storage: 10000
Quality control curve: X SD CV% L-J quality control curve for each test item
Flow cell: 32 ul quartz glass, 10mm
Lamp: Philips 6V 10W halogen lamp with longer life time
Screen: 7 inch color LCD
CPU: ARM series high speed micro CPU
Printer: Internal thermal sensitive printer
Interface: RS-232
Work Environment: Temperature 0°C~40°C, Humidity: 80%
Power supply: 100~240VAC, 50~60Hz, 100VA
Dimension: L340mm*W270mm*H160mm
Weight: 5kg
Liver Function: ALT AST ALP T-BIL TP ALB TTT CHE NH3
Renal Fucntion: BUN CRE UA URE CO2-CP Y-GT
Lipids: T-CHO TG HDL-C LDL-C
Cardiac function: CK CK-MB LDH
Immun proteins: IGA IGG IGM C3
Others: GLU HGB AMS
iCHEM 535
SEMI-AUTOMATED CHEMISTRY ANALYZER
Instrument: Fully-Auto Glycated Hemoglobin HbA1c Analyzer
Model: JL 760
Test Method: Low pressure ion exchange liquid chromatography (LPLC)
Test Item: HbA1ab HbA1c
Sample Type: Whole blood, hemolysis blood
Sample Volume: Whole blood 20ul, hemolysis blood 800ul
Photometer: 415nm LED Integral flow colorimeter
Analysis Time: 3.5mins
Precision: CV ≤ 3.0%
Display: 320*240 LC Graphic display, displaying real-time testing curve
Printer: 58nm Thermal printer, printing test curve and report
Panel: 0-9, YES, NO, START, STOP, ST, PRINT
Data Storage: 2G TF Card, more then 100,000 test reports storage (including Curve)
USB Port: Support remote upgrading, LIS Interface
PS-2 Port: External Barcode Scanner
Power Supply: 220V ± 22V, 50Hz ± 1Hz
Working Environment: Temperature 10°C~30°C, Relative humidity 30%~75%
Size: 550mm*405mm*347mm
Weight: 15kg
JL 760 (LPLC)
GLYCOSYLATED
HEMOGLOBIN
GLYCOSYLATED HEMOGLOBIN
Instrument: Fully-Auto Glycated Hemoglobin HbA1c Analyzer
Method: High pressure ion exchange liquid chromatography (HPLC)
Sampling System: Automatic identification system for Primary Tube & sample cup
Primary tube with cap puncture automatically
Throughput: <3mins per sample
Sample Tray: 15 samples position CAL (QC1) & CAL2 (QC2) position, ST
Sample Type: Venous Blood, Finger Peripheral Blood or hemolyse blood
Sample Volume: 5uL whole blood, or lyse blood (ratio 5uL:1000uL)
Test Range: HbA1c: 3-18%; 14-191mmol/mol, HbF:0.3-5%
Test Item: HbA1c HbF & B-Thalassemia
Heat system: Temperature range for column 5-40°c, can keep the constant
temperature, the fluctuation +-0.2°c
Light system: 415nm LED light, long life span, the fluctuation is 2mv per hour
Printer: 57mm Thermal printer, printing test curve and report
USB Port: Support LIS system, software upgrade online, external barcode scanner
optional
Barcode: External Barcode Scanner optional
Display: 8 inch LCD colorful touch screen, 162mm*122mm display real-time testing
curve
Data Storage: more than 10,000 test reports storage (including Curve), Inquiry freely
Language: English, Spanish, Russian, Portugese, Japanese
Report: NGSP, IFCC result
HbA1a, HbA1b, HbF, HbA1c, HbA0, variant sample report.
Optional PC software for regular report format
Work Condition: Temperature 5°C~40°C, Relative humidity 30%~75%
Dimension: 460mm*445mm*390mm